By Robert Berg
CHAPTER 22
22-1.1 Purpose. Recompression chambers are used for the treatment of decompression sickness, for surface decompression, and for administering pressure tests to prospective divers. Recompression chambers equipped for hyperbaric administra tion of oxygen are also used in medical facilities for hyperbaric treatment of carbon monoxide poisoning, gangrenous tissue, and other diseases. Decompres sion surface-supplied diving operations to depths greater than 130 fsw require that a chamber be available at the dive site.
22-1.2 Scope. This chapter will familiarize personnel with the maintenance and opera tional requirements for recompression chambers.
Most chamber-equipped U.S. Navy units will have one of five commonly provided chambers. They are:
22-2.1 Basic Requirements. Double-lock chambers are used because they permit tending personnel and supplies to enter and leave the chamber during treatment. Where stated:
�
On-site chamber is defined as a certified and ready chamber accessible withing 30 minutes of the dive site by available transportation.
�
On-station chamber is defined as a certified and ready chamber at the dive site.
�
Emergency chamber is defined as the closest recompression chamber available. A non-certified chamber may be used if the diving supervisor is of the opinion that it is safe to use.
Original Design Pressure – 200 psig Original Hydrostatic Test Pressure – 400 psig Maximum Operating Pressure – 100 psig
Figure 22-1. Double Lock Steel Recompression Chamber.
U.S. Navy Diving Manual—Volume 5
Design Pressure – 100 psig Original Hydrostatic Test Pressure – 200 psig Volume – Inner Lock = 136 cubic feet
– Outer Lock = 65 cubic feet
– Total = 201 cubic feet Principal locations – Repair/salvage ships and most shore-based facilities.
Figure 22-2. Double-Lock Aluminum Recompression Chamber.
Design Pressure – 100 psig Original Hydrostatic Pressure – 150 psig Principal Locations – ARS-50 Class Salvage Ships
Volume – Inner Lock = 134 cubic feet
Figure 22-3. ARS 50 Class Double Lock Recompression Chamber.
U.S. Navy Diving Manual—Volume 5
Figure 22-4. Fleet Modernized Double-Lock Recompression Chamber.
22-2.1.1 Chamber Volume. Navy chambers rated at the same pressure do not all have the same physical dimensions, with the exception of the aluminum chambers, ARS 50 class chambers, TRCS, and FARCC. Consequently, internal volumes of steel chambers are not standard and must be calculated for each chamber. Chamber volume is normally provided with the chamber.
The basic components of a recompression chamber are much the same from one model to another. They must be able to impose and maintain a pressure equivalent to a depth of 165 fsw (6 atmospheres absolute). The piping and valving on some chambers is arranged to permit control of the air supply and the exhaust from either the inside or the outside of the chamber. Controls on the outside must be able to override the inside controls in the event of a problem inside the chamber.
The usual method for providing this dual-control capability is through the use of two separate systems. The first, consisting of a supply line and an exhaust line, can only be controlled by valves that are outside of the chamber. The second air supply/exhaust system has a double set of valves, one inside and one outside the chamber. This arrangement permits the tender to regulate descent or ascent from within the chamber, but always subject to final control by outside personnel.
22-2.2 Modernized Chamber. Modernized chambers (Figure 22-4) have carbon dioxide and oxygen monitors, a CO2 scrubber system, a Built-In Breathing System (BIBS), and an oxygen dump system which together reduce the ventilation requirements. These chambers also include a chamber environment control system that regulates humidity and temperature.
22-2.3 Transportable Recompression Chamber System (TRCS). In addition to the chambers described above, a Transportable Recompression Chamber System (TRCS) is currently in fleet use (Figure 22-5). The TRCS consists of two pressure chambers. One is a conical-shaped chamber (Figure 22-6) called the Transportable Recompression Chamber, and the other is a cylindrical shaped vessel (Figure 22-7 ) called the Transfer Lock (TL). The two chambers are capable of being connected by means of a freely rotating NATO female flange coupling.
When a recompression chamber is required on site per Figure 6-14, or surface decompression dives are planned, the full TRCS system (including both TRC and TL) shall be on site.
When a recompression chamber is not required on site per Figure 6-14, the inner lock (TRC) may be used for emergency recompression treatment.
22-2.4 Fly Away Recompression Chamber (FARCC). This chamber system consists of a 60-inch double lock modernized chamber in a 20’ x 8’ x 8’ milvan (Figure 22-8 and Figure 22-9). The Fly Away Recompression Chamber (FARCC) also includes a life support skid (Figure 22-10). In addition, a stand-alone generator is provided for remote site power requirements.
U.S. Navy Diving Manual—Volume 5
| Height | 52″ with wheel, 48″ without wheels | ||
|---|---|---|---|
| Width | 50.7″ | ||
| Weight | 1,268 lbs. | ||
| Internal Volume | 45 cu. ft. | ||
| Door Opening | 26″ | ||
| View Ports | 3 @ 6″ dia. Clear Opening | ||
| Medical Lock | 5.75″ dia. x 11.8″ long | ||
| Mating Flange | Male per NATO STANG 1079 | ||
| Life Support Scrubber | Air driven, replaceable scrubber, canister | ||
| fits in Med Lock | |||
| BIBS | 2 masks – oxygen and air supply (with | ||
| capability for N2O2 or HeO2) – overboard | |||
| dump | |||
| Design Pressure | 110 psig | Atmospheric Monitoring | Oxygen and Carbon Dioxide Analyzer |
| Design Temperature | 0-125°F | Gas Supply | Primary and secondary air and O2 |
| Length | 95.7″ | Communications | Battery-powered speaker/headset phone |
| Furnishing | Patient litter, attendants seat | ||
22-2.5 Standard Features. Recompression chambers must be equipped with a means for delivering breathing oxygen to the personnel in the chamber. The inner lock should be provided with connections for demand-type oxygen inhalators. Oxygen can be furnished through a high-pressure manifold connected with supply cylin ders outside the chamber.
| Height | 52.9″ | ||
|---|---|---|---|
| Width | 54.8″ | ||
| Weight | 1,367 lbs. | ||
| Internal Volume | 45.5 cu. ft. | ||
| Door Opening | 2 doors – 26″ | ||
| View Ports | 2 @ 6″ dia. Clear Opening | ||
| Mating Flange | Rotating Female per NATO STANG | ||
| 1079 | |||
| Life Support Scrubber | Air-driven, replaceable scrubber, | ||
| canister fits in TRC Med Lock | |||
| BIBS | 2 masks – oxygen and air supply – | ||
| overboard dump | |||
| Design Pressure | 110 psig | Atmospheric Monitoring | Oxygen and Carbon Dioxide Analyzer |
| Design Temperature | 0-125°F | Gas Supply | Primary and secondary air and O2 |
| Length | 69.9″ | Communications | Sound-powered phone |
22-2.5.1 Labeling. All lines should be identified and labeled to indicate function, content and direction of flow. The color coding in Table 22-1 should be used.
U.S. Navy Diving Manual—Volume 5
22-2.5.2 Inlet and Exhaust Ports. Optimum chamber ventilation requires separation of the inlet and exhaust ports within the chamber. Exhaust ports must be provided with a guard device to prevent accidental injury when they are open.
22-2.5.3 Pressure Gauges. Chambers must be fitted with appropriate pressure gauges. These gauges, marked to read in feet of seawater (fsw), must be calibrated or compared as described in the applicable Planned Maintenance System (PMS) to ensure accuracy in accordance with the instructions in Chapter 4.
22-2.5.4 Relief Valves. Recompression chambers should be equipped with pressure relief valves in each manned lock. Chambers that do not have latches (dogs) on the doors are not required to have a relief valve on the outer lock. The relief valves shall be set in accordance with PMS. In addition, all chambers shall be equipped with a gag valve, located between the chamber pressure hull and each relief valve. This gag valve shall be a quick acting, ball-type valve, sized to be compatible with the relief valve and its supply piping. The gag valve shall be safety wired in the open position
22-2.5.5 Communications System. Chamber communications are provided through a diver’s intercommunication system, with the dual microphone/speaker unit in the chamber and the surface unit outside. The communication system should be arranged so that personnel inside the chamber need not interrupt their activities to operate the system. The backup communications system may be provided by a set of standard sound-powered telephones. The press-to-talk button on the set inside the chamber can be taped down, thus keeping the circuit open.
22-2.5.6 Lighting Fixtures. Consideration should be given to installation of a low-level lighting fixture (on a separate circuit), which can be used to relieve the patient of the heat and glare of the main lights. Emergency lights for both locks and an external control station are mandatory. No electrical equipment, other than that authorized within the scope of certification or as listed in the NAVSEA Autho rized for Navy Use (ANU) List, is allowed inside the chamber. Because of the possibility of fire or explosion when working in an oxygen or compressed air atmosphere, all electrical wiring and equipment used in a chamber shall meet required specifications.
Since a recompression chamber is emergency equipment, it must be kept in a state of readiness. The chamber shall be well maintained and equipped with all neces sary accessory equipment. A chamber is not to be used as a storage compartment. The chamber and the air and oxygen supply systems shall be checked prior to each use with the Predive Checklist and in accordance with PMS instructions. All diving personnel shall be trained in the operation of the recompression chamber equipment and should be able to perform any task required during treatment.
A recompression chamber system must have a primary and a secondary air supply system that satisfy the following requirements.
U.S. Navy Diving Manual—Volume 5
�
Primary. Sufficient air to pressurize the inner lock once to 165 feet and the outer lock twice to 165 feet and ventilate during one Treatment Table 4 (Chap ter 21).
�
Secondary. Sufficient air to pressurize the inner and outer locks once to 165 feet and ventilate for one hour at 70.4 scfm.
22-4.1 Capacity. Either system may consist of air banks and/or a suitable compressor. The primary recompression chamber support system must be capable of pressur izing the inner lock to a depth of 165 feet. The required total capacity is calculated as follows.
�
Primary System Capacity:
Cp = (5 × Vil) + (5 × Vol) + 4,224
Where:
CP = minimum capacity of primary system in scf = volume of inner lock in scf
Vil = volume of outer lock in scf
Vol 5 = atmospheres equivalent to 165 fsw 10 = twice 5 atmospheres 45,390 = total air in scf required to ventilate during a Table 4 Treatment
�
Secondary System Capacity:
Cs = (5 × Vil) + (5× Vol) + 4,224
Where:
Cs = minimum capacity of secondary system in scf = volume of inner lock
Vil = volume of outer lock
Vol
5 = atmospheres equivalent to 165 fsw
4224 = air in scf required for maximum ventilation rate of 70.4 scfm for one hour (60 min)
22-5.1 Predive Checklist. To ensure each item is operational and ready for use, perform the equipment checks listed in the Recompression Chamber Predive Checklist, Figure 22-11a .
�
Do not use oil on any oxygen fitting, air fitting, or piece of equipment.
�
Do not allow oxygen supply tanks to be depleted below 100 psig.
�
Ensure dogs are in good operating condition and seals are tight.
�
Do not leave doors dogged (if applicable) after pressurization.
�
Do not allow open flames, smoking materials, or any flammables to be carried into the chamber.
�
Do not permit electrical appliances to be used in the chamber unless listed in the Authorized for Navy Use (ANU).
�
Do not perform unauthorized repairs or modifications on the chamber support systems.
�
Do not permit products in the chamber that may contaminate or off-gas into the chamber atmosphere.
U.S. Navy Diving Manual—Volume 5
| RECOMPRESSION CHAMBER PREDIVE CHECKLIST | |
| Equipment | Initials |
| Chamber | |
| System certified | |
| Cleared of all extraneous equipment | |
| Clear of noxious odors | |
| Doors and seals undamaged, seals lubricated | |
| Pressure gauges calibrated/compared | |
| Air Supply System | |
| Primary and secondary air supply adequate | |
| One-valve supply: Valve closed | |
| Two-valve supply: Outside valve open, inside valve closed, if applicable | |
| Equalization valve closed, if applicable | |
| Supply regulator set at 250 psig or other appropriate pressure | |
| Fittings tight, filters clean, compressors fueled | |
| Exhaust System | |
| One-valve exhaust: valve closed and calibrated for ventilation | |
| Two-valve exhaust: outside valve open, inside valve closed, if applicable | |
| Oxygen Supply System | |
| Cylinders full, marked as BREATHING OXYGEN, cylinder valves open | |
| Replacement cylinders on hand | |
| Built in breathing system (BIBS) masks installed and tested | |
| Supply regulator set in accordance with OPs | |
| Fittings tight, gauges calibrated | |
| Oxygen manifold valves closed | |
| BIBS dump functioning |
Figure 22-11a. Recompression Chamber Predive Checklist (sheet 1 of 2). Figure 22-11b. Recompression Chamber Predive Checklist (sheet 2 of 2).
| RECOMPRESSION CHAMBER PREDIVE CHECKLIST | |
| Equipment | Initials |
| Electrical System | |
| Lights | |
| Carbon dioxide analyzer calibrated | |
| Oxygen analyzer calibrated | |
| Temperature indicator calibrated | |
| Carbon dioxide scrubber operational | |
| Chamber conditioning unit operational | |
| Direct Current (DC) power supply | |
| Ground Fault Interrupter (GFI) | |
| Communication System | |
| Primary system tested | |
| Secondary system tested | |
| Fire Prevention System | |
| Tank pressurized for chambers with installed fire suppression systems | |
| Combustible material in metal enclosure | |
| Fire-retardant clothing worn by all chamber occupants | |
| Fire-resistant mattresses and blankets in chamber | |
| Miscellaneous | |
| Inside Chamber: CO2-absorbent canister with fresh absorbent installed | |
| Urinal | |
| Primary medical kit | |
| Ear protection sound attenuators/aural protectors (1 set per person) | |
| Outside Chamber: Heater/chiller unit | |
| Stopwatches for recompression treatment time, decompression time, personnel leaving chamber time, and cumulative time | |
| Fresh CO2 scrubber canister | |
| U.S. Navy Diving Manual, Volume 5 | |
| Ventilation bill | |
| Chamber log | |
| Operating Procedures (OPs) and Emergency Procedures (EPs) | |
| Secondary medical kit | |
| Bedpan (to be locked in as required) |
U.S. Navy Diving Manual—Volume 5
22-5.3.1 Tender Change-Out. During extensive treatments, medical personnel may prefer to lock-in to examine the patient and then lock-out, rather than remain inside throughout the treatment. Inside tenders may tire and need relief.
22-5.3.2 Lock-In Operations. Personnel entering the chamber go into the outer lock and close and dog the door (if applicable). The outer lock should be pressurized at a rate controlled by their ability to equalize, but not to exceed 75 feet per minute. The outside tender shall record the time pressurization begins to determine the decompression schedule for the occupants when they are ready to leave the chamber. When the pressure levels in the outer and inner locks are equal, the inside door (which was undogged at the beginning of the treatment) should open.
22-5.3.3 Lock-Out Operations. To exit the chamber, the personnel again enter the outer lock and the inside tender closes and dogs the inner door (if so equipped). When ready to ascend, the Diving Supervisor is notified and the required decompression schedule is selected and executed. Constant communications are maintained with the inside tender to ensure that a seal has been made on the inner door. Outer lock depth is controlled throughout decompression by the outside tender.
22-5.3.4 Gag Valves. The actuating lever of the chamber gag valves shall be maintained in the open position at all times, during both normal chamber operations and when the chamber is secured. The gag valves must be closed only in the event of relief valve failure during chamber operation. Valves are to be lock-wired in the open position with light wire that can be easily broken when required. A WARNING plate, bearing the inscription shown below, shall be affixed to the chamber in the vicinity of each gag valve and shall be readily viewable by operating personnel.
The WARNING plates shall measure approximately 4 inches by 6 inches and read as follows:
WARNING The gag valve must remain open at all times. Close only if relief valve fails.
22-5.4 Ventilation. The basic rules for ventilation are presented below. These rules permit rapid computation of the cubic feet of air per minute (acfm) required under different conditions as measured at chamber pressure (the rules are designed to ensure that the effective concentration of carbon dioxide will not exceed 1.5 percent (11.4 mmHg) and that when oxygen is being used, the percentage of oxygen in the chamber will not exceed 25 percent).
22-5.4.1 Chamber Ventilation Bill. Knowing how much air must be used does not solve the ventilation problem unless there is some way to determine the volume of air actu ally being used for ventilation. The standard procedure is to open the exhaust valve a given number of turns (or fraction of a turn), which will provide a certain number of cubic feet of ventilation per minute at a specific chamber depth, and to use the supply valve to maintain a constant chamber depth during the ventilation period. Determination of valve settings required for different amounts of ventila tion at different depths is accomplished as follows.
U.S. Navy Diving Manual—Volume 5
WARNING This procedure is to be performed with an unmanned chamber to avoid exposing occupants to unnecessary risks.
a. Calculate how long it will take to change the chamber pressure by 10 feet if the exhaust valve lets air escape at the desired rate close to the depth in question. Use the following formula.
V60 ×Δ P
×
T = ------------------------------
R × (D +33 )
Where:
T = time in seconds for chamber pressure to change 10 feet V = internal volume of chamber (or of lock being used for test) in cubic feet (cf) R = rate of ventilation desired, in cubic feet per minute as measured at
chamber pressure (acfm) P = Change in chamber pressure in fsw D = depth in fsw (gauge)
Example: Determine how long it will take the pressure to drop from 170 to 160 feet in a 425-cubic-foot chamber if the exhaust valve is releasing 6 cubic feet of air per minute (measured at chamber pressure of 165 feet).
425 × 60 × 10
T = -------------------------------
6 165( + 33)
= 215 seconds
215 seconds
T = ------------------------------------------------
60 seconds / minute
= 3.6 minutes
b. Increase the empty chamber pressure to 5 feet beyond the depth in question. Open the exhaust valve and determine how long it takes to come up 10 feet (for example, if checking for a depth of 165 fsw, take chamber pressure to 170 feet and clock the time needed to reach 160 feet). Open the valve to different settings until you can determine what setting will approximate the desired time. Record the setting. Calculate the times for other rates and depths and determine the settings for these times in the same way. Make a chart or table of valve setting versus ventilation rate and prepare a ventilation bill, using this information and the ventilation rules.
�
The basic ventilation rules are not intended to limit ventilation. Generally, if air is reasonably plentiful, more air than specified should be used for comfort. This increase is desirable because it also further lowers the concentrations of carbon dioxide and oxygen.
�
There is seldom any danger of having too little oxygen in the chamber. Even with no ventilation and a high carbon dioxide level, the oxygen present would be ample for long periods of time.
�
These rules assume that there is good circulation of air in the chamber during ventilation. If circulation is poor, the rules may be inadequate. Locating the inlet near one end of the chamber and the outlet near the other end improves ventilation.
�
Coming up to the next stop reduces the standard cubic feet of gas in the cham ber and proportionally reduces the quantity (scfm) of air required for ventilation.
�
Continuous ventilation is the most efficient method of ventilation in terms of the amount of air required. However, it has the disadvantage of exposing the divers in the chamber to continuous noise. At the very high ventilation rates required for oxygen breathing, this noise can reach the level at which hearing loss becomes a hazard to the divers in the chamber. If high sound levels do occur, especially during exceptionally high ventilation rates, the chamber occupants must wear aural protectors (available as a stock item). A small hole should be drilled into the central cavity of the protector so that they do not pro duce a seal which can cause ear squeeze.
U.S. Navy Diving Manual—Volume 5
�
The size of the chamber does not influence the rate (acfm) of air required for ventilation.
�
Increasing depth increases the actual mass of air required for ventilation; but when the amount of air is expressed in volumes as measured at chamber pres sure, increasing depth does not change the number of actual cubic feet (acfm) required.
�
If high-pressure air banks are being used for the chamber supply, pressure changes in the cylinders can be used to check the amount of ventilation being provided.
22-6.1 Postdive Checklist. To ensure equipment receives proper postdive maintenance and is returned to operational readiness, perform the equipment checks listed in the Recompression Chamber Postdive Checklist, Figure 22-12a.
22-6.2 Scheduled Maintenance. Proper care of a recompression chamber requires both routine and periodic maintenance. Every USN recompression chamber (with the exception of the TRCS) shall be pressure tested upon installation, at 2-year inter vals thereafter, after a major overhaul or repair, and each time it is moved. This test shall be conducted in accordance with the pressure test for USN recompres sion chambers (Figure 22-13a) contained in this chapter. The completed test form shall be retained until retest is conducted. Chamber relief valves shall be tested in accordance with the Planned Maintenance System to verify setting. Each tested relief valve shall be tagged to indicate the valve set pressure, date of test, and testing activity. After every use or once a month, whichever comes first, the chamber shall receive routine maintenance in accordance with the Postdive Checklist. At this time, minor repairs shall be made and used supplies shall be restocked.
22-6.2.1 Inspections. At the discretion of the activity, but at least once a year, the chamber shall be inspected, both inside and outside. Any deposits of grease, dust, or other dirt shall be removed and, on steel chambers, the affected areas repainted.
22-6.2.2 Corrosion. Corrosion is removed best by hand or by using a scraper, being careful not to gouge or otherwise damage the base metal. The corroded area and a small area around it should then be cleaned to remove any remaining paint and/or corrosion.
22-6.2.3 Painting Steel Chambers. Steel chambers shall be painted in accordance with approved NAVSEA procedures. The following paint shall be utilized on steel chambers:
�
Inside:
| RECOMPRESSION CHAMBER POSTDIVE CHECKLIST | |
| Equipment | Initials |
| Air Supply | |
| All valves closed | |
| Air banks recharged, gauged, and pressure recorded | |
| Compressors fueled and maintained per technical manual/PMS requirements | |
| View Ports and Doors | |
| View-ports checked for damage; replaced as necessary | |
| Door seals checked, replaced as necessary | |
| Door seals lightly lubricated with approved lubricant | |
| Door dogs and dogging mechanism checked for proper operation and shaft seals for tight ness | |
| Chamber | |
| Inside wiped clean with Nonionic Detergent (NID) and warm fresh water | |
| All but necessary support items removed from chamber | |
| Blankets cleaned and replaced | |
| All flammable material in chamber encased in fire-resistant containers | |
| Primary medical kit restocked as required | |
| Chamber aired out | |
| Outer door closed | |
| CO2 canister packed | |
| Deckplates lifted, area below deckplates cleaned, deckplates reinstalled | |
| Support Items | |
| Stopwatches checked and reset | |
| U.S. Navy Diving Manual, Operating Procedures (OPs), Emergency Procedures (EPs), ven tilation bill and pencil available at control desk | |
| Secondary medical kit restocked as required and stowed | |
| Clothing cleaned and stowed | |
| All entries made in chamber log book | |
| Chamber log book stowed |
Figure 22-12a. Recompression Chamber Postdive Checklist (sheet 1 of 2).
U.S. Navy Diving Manual—Volume 5
| RECOMPRESSION CHAMBER POSTDIVE CHECKLIST | |
| Equipment | Initials |
| Oxygen Supply | |
| BIBS mask removed, cleaned per current PMS procedures, reinstalled | |
| All valves closed | |
| System bled | |
| Breathing oxygen cylinders fully pressurized | |
| Spare cylinders available | |
| System free of contamination | |
| Exhaust System | |
| One-valve exhaust: valves closed | |
| Two-valve exhaust: inside valves closed | |
| Two-valve exhaust: outside valves open | |
| All circuits checked | |
| Light bulbs replaced as necessary | |
| Pressure-proof housing of lights checked | |
| All power OFF | |
| Wiring checked for fraying |
Figure 22-12b. Recompression Chamber Postdive Checklist (sheet 2 of 2).
�
Outside:
22-6.2.4 Recompression Chamber Paint Process Instruction. Painting shall be kept to an absolute minimum. Only the coats prescribed above are to be applied. Naval Sea Systems Command will issue a Recompression Chamber Paint Process Instruction (NAVSEA-00C3-PI-001) on request.
22-6.2.5 Aluminum Chambers. Only steel chambers are painted. Aluminum chambers are normally a dull, uneven gray color and corrosion can be easily recognized. Aluminum chambers will not be painted.
22-6.2.6 Fire Hazard Prevention. The greatest single hazard in the use of a recompression chamber is from explosive fire. Fire may spread two to six times faster in a pres surized chamber than at atmospheric conditions because of the high partial
All U.S. Navy Standard recompression chambers are restricted to a maximum pressure of 100 psig, regardless of design pressure rating.
A pressure test shall be conducted on every USN recompression chamber (except TRCS):
Performance of the test and the test results are recorded on a standard U.S. Navy Recompres sion Chamber Air Pressure and Leak Test form (attached). The test is conducted as follows:
a. View-Port Leaks. Remove the view-port gasket (replace if necessary), wipe clean.
Acrylic view-ports should not be lubricated or come in contact with any lubricant. Acrylic view-ports should not come in contact with any volatile detergent or leak detector (non-ionic detergent is to be used for leak test). When reinstalling view-port, take up retaining ring bolts until the gasket just compresses evenly about the view-port. Do not overcompress the gasket.
b. Weldment Leaks. Contact appropriate NAVSEA technical authority for guidance on corrective action.
Figure 22-13a. Pressure Test for USN Recompression Chambers (sheet 1 of 3).
U.S. Navy Diving Manual—Volume 5
Ship/Platform/Facility _____________________________________________________________ Type of Chamber: Double-Lock Aluminum
Double-Lock Steel
Portable Recompression Chamber
Other* ___________________________________________________
Manufacturer ___________________________________________________________________ Date of Manufacture______________________________________________________________ Contract/Drawing No._____________________________________________________________ Maximum Working Pressure _______________________________________________________ Date of Last Pressure Test ________________________________________________________ Test Conducted by _______________________________________________________________
(Name/Rank)
1. Conduct visual inspection of chamber to determine if ready for test Chamber Satisfactory ______________ Initials of Test Conductor ____________________ Discrepancies from fully inoperative chamber equipment:
Close inner door lock. With outer lock door open pressure inner lock to 100 fsw (45 psig) and verify that the following components do not leak:
(Note: If chamber has medical lock, open inner door and close and secure outer door.)
Inner lock leak checks Initials of Test Conductor
A. Shell penetrations and fittings _____________________ Satisfactory
B. View Ports _____________________ Satisfactory
C. Door Seals _____________________ Satisfactory
D. Door Dog Shaft Seals _____________________ Satisfactory
E. Valve Connections and Stems _____________________ Satisfactory
F. Pipe Joints _____________________ Satisfactory
G. Shell Welds _____________________ Satisfactory
3. Increase inner lock pressure to 225 fsw (100 psig) and hold for 5 minutes. Record Test Pressure ______________________ Satisfactory ________________________
(Note: Disregard small leaks at this pressure).
Figure 22-13b. Pressure Test for USN Recompression Chambers (sheet 2 of 3).
4. Depressurize lock slowly to 165 fsw (73.4 psig). Secure all supply and exhaust valves and hold for one hour.
Start Time ___________________________ Pressure 165 fsw
End Time ___________________________ Pressure ________________ fsw If pressure drops below 145 fsw (65 psig) locate and mark leaks. Depressurize, repair, and retest inner lock. Inner Lock Pressure drop test passed _______________ Satisfactory Initials of Test Conductor.
Test Director Date
Diving Officer Date
Commanding Officer Date
Figure 22-13c. Pressure Test for USN Recompression Chambers (sheet 3 of 3).
U.S. Navy Diving Manual—Volume 5
pressure of oxygen in the chamber atmosphere. The following precautions shall be taken to minimize fire hazard:
� Maintain the chamber oxygen percentage as close to 21 percent as possible and never allow oxygen percentage to exceed 25 percent.
� Remove any fittings or equipment that do not conform with the standard requirements for the electrical system or that are made of flammable materials. Permit no wooden deck gratings, benches, or shelving in the chamber.
� Use only mattresses designed for hyperbaric chambers. Use Durett Product or submarine mattress (NSN 7210-00-275-5878 or 5874). Other mattresses may cause atmospheric contamination. Mattresses should be enclosed in flame proof covers. Use 100% cotton sheets and pillow cases. Put no more bedding in a chamber than is necessary for the comfort of the patient. Never use blan kets of wool or synthetic fibers because of the possibility of sparks from static electricity.
� Keep oil and volatile materials out of the chamber. If any have been used, ensure that the chamber is thoroughly ventilated before pressurization. Do not put oil on or in any fittings or high-pressure line. If oil is spilled in the cham ber or soaked into any chamber surface or equipment, it must be completely removed. If lubricants are required, use only those approved and listed in Naval Ships Technical Manual (NSTM) NAVSEA S9086-H7-STM-000, Chapter 262. Regularly inspect and clean air filters and accumulators in the air supply lines to protect against the introduction of oil or other vapors into the chamber. Permit no one to wear oily clothing into the chamber.
� Permit no one to carry smoking materials, matches, lighters or any flammable materials into a chamber. A WARNING sign should be posted outside the chamber. Example:
WARNING Fire/Explosion Hazard. No matches, lighters, electrical appliances, or flammable materials permitted in chamber.
22-6.2.6.1 Fire Extinguishers. Only fire extinguishers listed on the NAVSEA Authorized for Navy Use (ANU) Lists are to be used.
All U.S. Navy diver candidates shall be physically qualified in accordance with the Manual of the Medical Department, Art. 15-66. Candidates shall also pass a pressure test before they are eligible for diver training. This test may be conducted at any Navy certified recompression chamber, provided it is administered by qual ified chamber personnel.
22-7.1 Candidate Requirements. The candidate must demonstrate the ability to equalize pressure in both ears to a depth of 60 fsw. The candidate shall have also passed the screening physical readiness test in accordance with MILPERSMAN 1410380, Exhibit 5.
22-7.2 Inside Tender Requirements. The inside tender(s) should be a qualified diver.
22-7.3.1 References.
�
Navy Military Personnel Manual, Art. 1410380
�
Manual of the Medical Department, Art. 15-66
�
SECNAVINST 12000.2A